DISTRITRIP (cutie de 50 de seringi) |
Parte alicota |
Cod catalog |
Volum (μL) |
Erori maxime permise Gilson | Erori maxime permise ISO 8655 | ||
|
Eroare sistematica µL |
Eroare aleatoare µL |
Eroare sistematica µL |
Eroare aleatoare µL |
||||
|
125 µL Micro Micro ST |
1µL pana la 12.5 µL |
F164100 F164130 |
Min: 2 5 Max:10 |
±0.100 ±0.125 ±0.200 |
≤0.080 ≤0.075 ≤0.200 |
±0.2 ±0.2 ±0.2 |
≤0.1 ≤0.1 ≤0.1 |
|
1250 µL Mini Mini ST |
10µL pana la 125 µL |
F164110 F164140 |
Min: 20 50 Max.100 |
±0.80 ±1.00 ±1.00 |
≤0.20 ≤0.40 ≤0.60 |
±1.5 ±1.5 ±1.5 |
≤0.6 ≤0.6 ≤0.6 |
|
12.5 mL Maxi Maxi ST |
100µL pana la 1.25 mL |
F164120 F164150 |
Min. 200 500 Max. 1000 |
±6.0 ±7.5 ±10.0 |
≤1.0 ≤1.5 ≤2.5 |
±12.0 ±12.0 ±12.0 |
≤4.0 ≤4.0 ≤4.0 |
| Cod producator | Descriere | Cod SEPADIN |
| F164001 | DISTRIMAN pipeta mecanica repetitiva | 327-020 |
| F164100 | DISTRITIP 125 µL Micro | 327-165 |
| F164130 | DISTRITIP 125 µL Micro ST | 327-276 |
| F164110 | DISTRITIP 1250 µL Mini | 327-021 |
| F164140 | DISTRITIP 1250 µL Mini ST | 327-277 |
| F164120 | DISTRITIP 12.5 mL Maxi | 327-166 |
| F164150 | DISTRITIP 12.5 mL Maxi ST | 327-278 |